(Created page with "{{stub}} {{Repair Guide |Device= |Affects parts= |Needs equipment= |Type= |Difficulty= }} ==Problem description== Fixing the ChargeMaster Medium that doesn't turn on, does not pull any current, and does not react to power. This particular issue is related to the main DC-DC transformer being faulty. This is a rather common issue with this device since the DC-DC circuit does not use any smart IC to reglaute it but rather an analog circuit to drive the switching MOSFETs tha...")
 
No edit summary
 
Line 1: Line 1:
{{stub}}
{{Repair Guide
{{Repair Guide
|Device=
|Device=Simco ChargeMaster Medium
|Affects parts=
|Affects parts=Main switching transformer
|Needs equipment=
|Needs equipment=multimeter, soldering iron, soldering station
|Type=
|Type=Soldering, Teardown, Part replacement
|Difficulty=
|Difficulty=3. Hard
}}
}}
{{stub}}
==Problem description==
==Problem description==
Fixing the ChargeMaster Medium that doesn't turn on, does not pull any current, and does not react to power. This particular issue is related to the main DC-DC transformer being faulty. This is a rather common issue with this device since the DC-DC circuit does not use any smart IC to reglaute it but rather an analog circuit to drive the switching MOSFETs that switch the transformer.
Fixing the ChargeMaster Medium that doesn't turn on, does not pull any current, and does not react to power. This particular issue is related to the main DC-DC transformer being faulty. This is a rather common issue with this device since the DC-DC circuit does not use any smart IC to reglaute it but rather an analog circuit to drive the switching MOSFETs that switch the transformer.

Latest revision as of 07:29, 30 June 2024

Simco ChargeMaster Medium not turning on repair
Device Simco ChargeMaster Medium
Affects part(s) Main switching transformer
Needs equipment multimeter, soldering iron, soldering station
Difficulty ◉◉◉◌ Hard
Type Soldering, Teardown, Part replacement


This article is a stub. You can help Repair Wiki grow by expanding it

Problem description

Fixing the ChargeMaster Medium that doesn't turn on, does not pull any current, and does not react to power. This particular issue is related to the main DC-DC transformer being faulty. This is a rather common issue with this device since the DC-DC circuit does not use any smart IC to reglaute it but rather an analog circuit to drive the switching MOSFETs that switch the transformer.

Main power board, transformer on the top right (Figure 1)

Symptoms

  • Doesn't turn on.
  • No reaction to input AC power
  • Screen always off

Solution

This issue was rather hard to diagnose since there aren't any ICs that you can check the various pins of. After checking if the MOSFETs are not shorted, the resistors/diodes/caps are within spec, the only thing left to check was the transformer. There's currently no way to buy a replacement transformer, the only way is to open it up, unwind the wires making note of the number of turns, and then replace the windings with a slightly thicker wire. This problem occurs because the transformer heats up beyond its rated specification which cause the enamel on the windings to break and short them.

If the issue has a single, specific solution, provide it in this section and delete the sub-headers below. Otherwise, if there are multiple potential causes for the problem, outline the process of identifying the root cause and offer corresponding solutions below.

Diagnostic Steps

Repair Steps